TESTING OF PULP VITALITY BY PULSOXIMETRY

Authors: Gabriela CIOBANU, Ileana ION, Luiza UNGUREANU

Abstract:

The methods applied for diagnosing the health condition of the pulp tissue are numerous, however, nowadays, an increasingly higher number of conventional tests are replaced by some objective, non-invasive, painless and reliable tests. Among them, pulse oximetry is a method for the investigation of pulp vitality based on oxygen saturation (SaO2) of the hemoglobin from the blood present in the pulp vascular bed, as a means of differentiating among the vital and the non-vital teeth. In the present study, registrations were made on a group of 120 frontal maxillary teeth, in patients with ages between 20 and 40 years, on using a digital sensor modified by the pulse oximeter with which the pulse and the values of oxygen saturation were measured at the level of both teeth and right hand finger. The mean SaO2 value in the pulp blood of the vital teeth was of 83.30% for the central incisor, of 78.51% for the lateral one and of 84.56%, respectively, for the canine; the value recorded at finger level was of 97%. In the non-vital teeth, the SaO2 value measured on the pulse oximeter was of 0%. Pulse registration showed mean values of 70.56 beatings/min at tooth level and of 70.88 beatings/min, respectively, at finger level.
The results of the present study may confirm that pulse oximetry represents a simple, non-traumatic, efficient and objective method for testing the vitality condition of the dental pulp.

Keywords:
  • endodontic diagnosis
  • pulse oximeter
  • tests of pulp vitality.